New Strain Gage Version is Available
|
|
Built-In USB Programming Port
|
|
Modular Output Modules
|
|
Universal Process, Voltage, Current, Resistance and Temperature Inputs
|
|
Universal AC/DC Power Supply
|
|
6/9 Digit Dual Line/ Tri-Color Display with 18 mm (0.71") and 8.9 mm (0.35")
|
|
Programmable Units Display
|
|
Variable Contrast and Intensity Display
|
|
Up to 160 Samples Per Second Conversion Rate
|
|
NEMA 4X (IP65) Sealed Front Bezel
|
View related products - Instrumentation for Pressure and Strain Measurements
|
| |
|
|
|
The DP63100 analog process panel
meter offers many features and
performance capabilities to suit a
wide range of industrial applications.
The DP63100 has a universal input
to handle various input signals
including DC voltage/current,
process, resistance, strain and
temperature. The optional plug-in
output cards allow the opportunity to
configure the meter for present
applications, while providing easy
upgrades for future needs. The
DP63100 employs a dual line,
tri-color display with a large 18 mm
(0.71"), tri-color 6 digit top display
line and a 8.9 mm (0.35"), 9 digit
green bottom display line. The
meter provides a MAX and MIN
reading memory with programmable
capture time. The capture time is
used to prevent detection of false
max or min readings which may
occur during start-up or unusual
process events.The signal totalizer
(integrator) can be used to compute
a time-input product. This can be
used to provide a readout of
totalized flow or calculate service
intervals of motors, pumps, etc.
The meter has up to four setpoint
outputs, implemented on plug-in
option cards. The plug-in cards
provide dual FORM-C relays, quad
FORM-A, or either quad sinking or
quad sourcing open collector logic
outputs. The setpoint alarms can be
configured to suit a variety of control
and alarm requirements.
Communication and bus capabilities
are also available as option cards.
These include RS232 and RS485.
The DP63100 can be programmed
to utilize MODBUS® protocol. With
MODBUS, the user has access to
most configuration parameters.
Readout values and setpoint alarm
values can be controlled through
the bus. Additionally, the meter has
a feature that allows a remote
computer to directly control the
outputs of the meter. The DP63100
includes a built-in USB
programming port. With a Windows®
based program, configuration data
can be downloaded to the unit
without the need of any additional
option cards. A linear DC output
signal is available as an optional
plug-in card. The card provides
either 20 mA or 10V signals. The
output can be scaled independent
of the input range and can track
either the input, totalizer, max or
min readings. The meter has been
specifically designed for harsh
industrial environments. With
NEMA 4X (IP65) sealed bezel and
extensive testing of noise effects
and CE requirements, the meter
provides a tough reliable
application solution.
SPECIFICATIONS
Display: Negative image LCD
Top Line: 6 digit, 18 mm (0.71"), with tri-color backlight (red, green or orange)
Display Range: -199999 to 999999;
Bottom Line: 9 digit, 8.9 mm (0.35"), with green backlight
Display Range: -199,999,999 to 999,999,999
Power
AC Power: 50 to 250 Vac, 50/60 Hz, 14 VA
DC Power: 21.6 to 250 Vdc, 8 W
Isolation: 2300 Vrms for 1 min to all inputs and outputs. KEYPAD: 3 programmable function keys, 5 keys total
A/D Converter: 24 bit resolution
Update Rates: A/D conversion rate; programmable 5 to 160 readings/sec
Thermocouple Inputs
Readout:
Scale: °F or °C
Offset Range: -199,999 to 999,999 display units
Input Impedance: 20 MO
Lead Resistance Effect: 0.03 µV/Ω
Max Continuous Overvoltage: 30V
RTD Inputs
Type: 3 or 4 wire, 2 wire can be compensated for lead wire resistance
Excitation Current: 100 Ω
Range: 136.5 µA ±10%
10 Ω Range: 2.05 mA ±10%
Lead Resistance: 100 Ω
Range: 10 Ω/lead max
10 Ω Range: 3 Ω/lead max
Max Continuous Overload: 30V
Excitation Power: Jumper selectable
Transmitter Power: ±18 Vdc @ 50 mA
Reference Voltage: ±2 Vdc, ±2%
Strain Gage Inputs
Connection Type: 4-wire bridge (differential); 2-wire (single-ended)
Common Mode Range (with respect to input common): 0 to +5 VDC
Rejection: 80 dB (DC to 120 Hz)
Input Ranges:
± 24 mVDC
± 240 mVDC
User Inputs
Two Programmable User Inputs Max Continuous Input: 30 Vdc
Isolation to Sensor Input Common: Not isolated
Response Time: 12 msec max
Totalizer
Time Base: Second, minute, hour, or day
Batch: Can accumulate (gate) input display from a user input
Time Accuracy: 0.01% typical
Decimal Point: 0 to 0.0000
Scale Factor: 0.001 to 65.000
Low Signal Cut-Out: -199,999 to 999,999
Total: 6 digits on line 1; 9 digits on line 2
Custom Linearization
Data Point Pairs: Selectable from 2 to 16
Display Range: -199,999 to 999,999
Decimal Point: 0 to 0.0000
Memory: Nonvolatile FRAM memory retains all programmable parameters and display values
Environmental Conditions
Operating Temperature Range: 0 to 50°C (32 to 122°F) 0 to 45°C (32 to 113°F) with all three plug-in cards installed
Storage Temperature Range: -40 to 60°C (-40 to 140°F)
Operating and Storage Humidity: 0 to 85% max RH non-condensing
Altitude: Up to 2000 m (6561')
Connections: High compression age-clamp terminal block
Construction: This unit is rated for NEMA 4X (IP65) outdoor use IP20 touch safe. Installation Category II, pollution degree 2. One piece bezel/case. Flame resistant. Synthetic rubber keypad. Panel gasket and mounting clip included.
Dimensions
Case: 44.45 H x 91.44 W x 105 mm D (1.75 x 3.6 x 4.14")
Bezel: 49.53 x 96.52 mm (1.95 x 3.8")
Panel Cutout: 1/8 DIN, 45 x 92 mm (1.77 x 3.62")
Weight: 226.8 g (8 oz)
|
| |
|
|
|
|
|
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with operator’s manual. Adding option cards to meters can be fitted with up to three optional plug-in cards. However, only one card from each function type can be installed at a time. The function types include Setpoint Alarms, Communications, and Analog Output. The cards can be installed initially or at a later date. Each optional plug-in card is shipped with installation and programming instructions.
|
|
|
|
|
| |
Related Links -
Manual
|
Product Manual |
Download DP63100 - Analog Input Panel Meter |
Download Strain Gage Input Panel Meter |