Relative Humidity/Ambient Temperature Transmitters
|
Click for larger image. |
|
|
USD$14,100.00
HX94BV0W
|
|
RoHS Certified
|
|
Accurate to ±2.5% Relative Humidity and ±0.6°C (±1°F) Temperature
|
|
Two Separate Analog Outputs, Humidity and Temperature
|
|
Four Analog Outputs available, 4 to 20 mA, 0 to 1V, 0 to 5V, 0 to 10V
|
|
Stainless Steel NEMA 4 (IP66) Enclosure
|
|
Three Different Output Connection Styles Available
|
|
Includes Hardware for Wall and Duct Mounting
|
|
Removable Stainless Steel Sensor Cap
|
View related products - Relative Humidity Measurement Instruments
|
| |
|
|
|
The OMEGA™ HX94B Series
temperature/relative humidity
transmitter provides a linearized and
temperature compensated output
signal of 4 to 20 mA, 0 to 1 Vdc,
0 to 5 Vdc or 0 to 10 Vdc depending
upon the model selected for both
temperature/relative humidity
measurement. The output signals
have been calibrated and scaled to
0 to 100°C for temperature and
0 to 100% for relative humidity.
The digital temperature and relative
humidity sensor is protected by a
stainless steel cap that is easily
removed for cleaning. The NEMA 4
stainless steel enclosure and cable
entry connection provides weather
tight protection. The stainless steel
enclosure terminates to a 6 feet
shielded cable. The cable end is
either stripped leads or terminates
to a 4-pin connector.
Specifications
Temperature
Measuring Range: 0 to 100°C (32°F to 212°F)
Accuracy: ±0.6°C from 0 to 50°C (32 to 122°F), ±1.25°C from 50 to 100°C (122 to 212°F), for all models except:
HX94BC,HX94BCNPT,HX94BCW: ±1.2°C from 0 to 50°C (32 to 122°F); ±1.9°C from -30 to 0°C (-22 to 32°F) and 50 to 75°C (122 to 167°F)
Repeatability: ±0.2°C (0.4°F)
Response Time: 5 seconds minimum, 30 seconds maximum
Sample Rate: 1 sample every 4 seconds
Relative Humidity
Measuring Range: 0 to 100% RH
Accuracy: ±2.5 % from 20 to 80% RH, ±3.5% from 5 to 20% and 80 to 95% RH, ±4% from 0 to 5% and 95 to 100% RH
Hysteresis: ±1%RH
Repeatability: ±0.1%RH
Response Time: 8 seconds typical
Sample Rate: 1 sample every 4 seconds
Operating Temperature Range: 0 to 100°C (32 to 212°F)
Analog Outputs for Temperature and Humidity: 4 to 20 mA, 0 to 1V, 0 to 5V,
0 to 10V
Input Power Supply: 9 to 30 Vdc @ 100 mA: 4 to 20 mA, 0 to 1V, 0 to 5V output,
12 to 30 Vdc @ 100 mA: 0 to 10V output
Max Current Loop Resistance: Ω = [(V supply - 4V)/0.02A)] -50; 200Ω @ 9V supply or 1250Ω @ 30V supply
Max Load Resistance: 1.25K (for all voltage outputs)
Sensor Type: Digital sensor
Enclosure Housing: 316 stainless steel, NEMA 4; non-NPT version includes IP68-rated cable gland
Electrical Connections:
HX94B**W, HX94B**NPT: 1.8 m (6') 4 conductor PTFE shielded
cable with stripped leads termination
HX94B**: 1.8 m (6') 4 conductor PTFE shielded cable with a 4-pin connector termination; a 4-pin mating connector is included that accepts 18 to 26 AWG wire
Dimensions: 258 L x ø19 mm OD (10.15 x 0.75")
Weight: 200 g (7 oz) with mounting kit
|
| |
|
|
|
|
|
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with mating connector (models HX94BC, HX94BV0, HX94BV1, HX94BV2 only), removable stainless steel cap, removable duct flange, wall mounting kit, and quick start manual.
Ordering Example: (1) HX94BC Temperautre/ Relative Humidity Transmitter with 4-20mA output terminated to 4 pin connector, USD$15,640.00
|
|
|
|
|
| |
|
Back to top
|
|
|