USD$207,355.00
OM-CP-AVS140-6
|
|
Complete Systems for Autoclave Validation
|
|
Enables Compliance with FDA 21 CFR Part 11
|
|
NIST Traceable Calibration Included
|
|
OM-CP-AVS140-1 for Temperature Monitoring
|
|
OM-CP-AVS140-6 for Temperature and Pressure Monitoring
|
View related products - Data Loggers
|
| |
|
|
|
The OM-CP-AVS140-1 and OM-CP-AVS140-6
are complete systems used to perform autoclave
validations. The OM-CP-AVS140-1 system for
autoclave temperature monitoring consists of a NIST
traceable OM-CP-HITEMP-140-1 temperature data
logger, an OM-CP-IFC400 USB docking station and
FDA 21 CFR part 11 secure software. For larger size
autoclaves or mapping projects, the OM-CP-AVS140-6
is a system for autoclave temperature and pressure
monitoring that includes five OM-CP-HITEMP-140-1
temperature data loggers, one OM-CP-PR140 pressure
data logger, an OM-CP-IFC406 multiplexer interface
(allows up to 6 data loggers to be interfaced to a single
USB port on the computer) and the secure software.
The secure software aids users in compliance with
FDA 21 CFR Part 11 requirements. The software
ensures standards in which electronic files are
considered equivalent to paper records, saving time
and effort. The secure software contains criteria such
as electronic signatures, access codes, secure data
files and an audit trail which meets the requirements
of 21 CFR Part 11 and helps provide data integrity.
IQ/OQ/PQ (Installation Qualification/Operational
Qualification/Performance Qualification) protocols
are included to validate that the software has been
installed and is operating correctly. The layout of the
secure software is similar to the OM-CP Series Data
Logger standard software, allowing users to easily
learn the additional features. The Windows® based
software package allows the user to effortlessly
collect, display and analyze data. A variety of powerful
tools provide the ability to examine, export, and print
professional looking data with just a click of the mouse.
SPECIFICATIONS
OM-CP-HITEMP140-1 Temperature Data Logger with 27 mm (1.1") Probe
Temperature Sensor: 100Ω Platinum RTD
Temperature Range (Body): -40 to 140°C (-40 to 284°F)
Temperature Measurement Range (Probe): -200 to 260°C (-328 to 500°F)
Temperature Resolution: 0.01°C (0.02°F)
Calibrated Accuracy: ±0.1°C (±0.18°F) [20 to 140°C (68 to 284°F)]
Battery Life: 1 year typical [1 minute reading rate at 25°C (77°F)]
Data Format: Date and time stamped °C, °F, K, °R
Time Accuracy: ±1 minute/month at 20 to 30°C (68 to 86°F) (stand alone mode)
Operating Environment: -40 to 140°C (-40 to 284°F), 0 to 100% RH
Dimensions
Body: 40 H x 25 mm D (1.6 x 0.97")
Probe: 27 L x 3.2 mm dia (4.8 mm transitional dia) [1.1 x 0.125" dia (0.188" transitional dia)]
Weight: 120 g (4.2 oz)
Material: 316 stainless steel
OM-CP-PR140 PRESSURE DATA LOGGER
Pressure Sensor: Semiconductor strain gauge
Pressure Range: 0 to 5 bar (0 to 72.5 psia)
Pressure Resolution: 0.1 mBar (0.0015 psia)
Calibrated Accuracy: ±0.03 Bar (±0.435 psi) (20 to 140°C/68 to 284°F)
Pressure Response Time: 0.1 ms (10 to 90% FSR)
Repeatability: ±0.5% FSR; ±0.2% typical
Data Format: psia, bar, mbar, atm, Torr, Pa, kPa, MPa, mmHg, inHg, inH20, mH20, ft, m
Battery Life: 2 years typical
Time Accuracy: ±1 minute/month at 20°C
Operating Environment: -20 to 140°C, (-4 to 284°F) 0 to 100% RH, may be used above 60°C (140°F) for up to 8 hrs per 24 hr period
Dimensions: 50.8 L x 25.4 mm dia. (2.0 x 1.0")
Weight: 85 g (3 oz)
Enclosure: Stainless steel
COMMON SPECIFICATIONS (OM-CP-HITEMP140-1 and OM-CP-PR140)
Real Time Recording: May be used with PC to monitor and record data in real time
Memory: 32,700 readings; software configurable memory wrap
Reading Rate: One second up to once every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6V high-temperature lithium battery (included)
Software: XP SP3/Vista/7 and 8 (32-bit and 64-bit)
Computer Interface: OM-CP-IFC400 USB docking station or OM-CP-IFC406 multiplexer interface required, 125,000 baud
|
| |
|
|
|
|
|
† All dollar amounts on this site are shown in US currency.
Note: OM-CP-HITEMP140-1 temperature data loggers and OM-CP-PR140 pressure data loggers come complete with 3.6V high temperature lithium battery.
Ordering Example: (1) OM-CP-AVS140-1 Autoclave temperature validation system: includes one (1) OM-CP-HITEMP-140-1 temperature data logger with NIST calibration certificate, OM-CP-IFC400 USB docking station and FDA 21 CFR Part 11 Secure Software with IQ/OQ/PQ validation protocols, USD$72,450.00
|
|
|
|
|
| |
Related Links -
Manual, Related Products
|
Product Manual |
Download High Temperature Data Loggers Quick Start Guide |
Download Autoclave Validation Data Logging System |
Related Products |
|
|
|
|
|
|
|
|